|More
Elisa
Readers|Multiskan
Readers|Strip
Plate Washers|Great Deals on Washers|
Great Deals
Microplate
Readers 2 Elisa Plate Readers
Pages 4,
5, 6
Elisa Plate
Readers...Overstocks,
demos and remanufactured instruments. Instruments are supported with six month to three year warranty
and 6 days a week technical support. In stock items include
New, demo and
remanufactured Monochromatic, 96 Well or 96/384 Well Elisa Plate Readers
to include Molecular Devices, Labsystems, Thermo Electron, Dynex, Biorad and Bio-tek
Readers (different models, please inquire).

Bio-tek µQuant™
Monochromatic Microplate Spectrophometer.
Spectral
Monochromatic Wavelength Range 200-999nm in 1nm increments. Reads 6, 12, 24,
48, 96, 384 well plates. Endpoint, single
and dual spectral range
reads, kinetic reads, spectral skans and well area scanning. Absorbance measurement range 0-4.000 OD, ±2
nm accuracy, ±0.2nm repeatability, 2.4nm bandpass. Extensive
onboard data analysis and storage for DNA quantitation and 260/280 ratio, endpoint
calculations, curve fitting options, more features included with PC software.
Self testing operations include system test, checksum test, absorbance plate
test (NIST) and liquid testing. Small footprint, huge features,
NIST (National Institutes Standards & Technology) certification
and documentation provided.
One year warranty on all remanufactured models. Newest
model sells new for $14,000.
|
Catalog # |
Instrument Description/Price/# in stock |
|
P97239 |
µQuant Reader
Remanufactured/New $5,700. (2) |
|
888777 |
KC Jr. software |
|
888999 |
Gen 5 PC Software |
Tecan
Sunrise Microplate Readers. Spectral Wavelength Range 340-750nm, with four
standard
filters.
340 to 399nm: 0 to 3.0 OD, 400 to 750nm: 0
to 4.0 OD.
Or, purchase addl.
filter slide for gradient 1nm increment readings. Can be used as
Agglutination Reader with added software. Designed to read a
96-well plate in 6 seconds in dual wavelength mode (8 seconds) with accuracy and
linearity better than 1% and precision within 0.5%. 12 measurement
channels and 1 reference channel.
Linear shaking in 4 different
modes. Endpoint, kinetic assays read time 5 seconds
with capability for excel formatting. *New model has onboard screen
for stand alone use with added software.
Small footprint,
easy to use, huge features, NIST (National Institutes Standards &
Technology) certification and documentation provided
on each filter wavelength.
(Touchscreen on new model).
One year warranty on all models.
|
Catalog # |
Instrument Description/Price/# in stock |
|
P97269 |
Tecan Sunrise Reader *NEW (1) |
|
P97259 |
Tecan Sunrise Reader Remanufactured
(2) |
|
Magellan |
Magellan PC Software |
EL
311, 311s, 340 96 well Microplate Readers.
Spectral Range
400-750nm wavelength (model EL 340 range 340-750nm) with 5 second read
and 3 filters included. Kinetic/endpoint, single and dual
wavelength reads. Shaking at low, mid and high levels. Absorbance range 0.000 to 4.000 OD with narrow
bandwidth filters as listed below included. Full spectrum of operation with PC
software use.
Models come with
NIST (National Institutes Standards &
Technology) certification
and documentation provided
calibrated on each filter wavelength. Turnkey systems to interface with printer or
PC software for full operation.
One year warranty
on all remanufactured models.
|
Catalog # |
Instrument Description/Price/# in stock |
|
P97267 |
EL311 with 405, 450, 490nm filters $2,295. (2) |
|
P97248 |
EL311s with 405, 450nm filters $2,295. (2) |
|
P97268 |
EL340 shown with 405, 450nm filters $2,295. (2) |
|
99866 |
Inkjet
Printer |
|
6723121 |
Co Inkjet Printer |
iEMS
96 well Microplate Reader. Spectral Range 340-850nm 96 well
Microplate based instrument.
Upper and lower plates provide built-in temperature control and prevent
plate condensation. With shaking capabilities and wide
measurement range of up to 4.000 OD makes it an ideal instrument for
Microbiological studies. Wavelength range
340-850nm and comes with user's
choice of 2 filters in an eight place filter wheel, NIST (National
Institutes Standards & Technology)
certification and documentation provided on each filter wavelength.
Three year warranty on all demo/remanufactured models.
Turnkey
system.
100-240 V.
|
Specifications |
| light source |
tungsten-halogen lamp 8V/50W |
| type of
filters |
8 position
filter wheel with installed filters: 340, 405, 414, 450, 492, 540, 620, 690nm
(2 of choice) |
| wavelength
accuracy |
2.0 nm stand
alone 1.00nm with PC software |
| wavelength
range |
340-850nm |
| linearity |
0...4.000 A
max. dev. less than 2% 0...3.000 A
at 340 nm and 620 nm |
| dynamic range |
0.000 A ...
6.000 A |
| resolution |
0.001 A |
| accuracy |
1% or 0.002
A, whichever is greater (0...2.000 A), 2% (2.000...3.000 A) |
| precision |
CV<0.2% or
0.002A, whichever is greater (0...3.000A) CV<1.0% at 4.000A |
| Measurement
speed |
6 seconds on
96 well plate |
|