Perkin Elmer Victor 2 Fluorescence, Luminescence, and
Absorbance Reader

Description:

VICTOR 2 multi-task plate reader designed to support the future demands of industrial and academic laboratories for multiple assay technologies on a single platform.   Top of the line plate readers for fluorescence,  luminescence, and absorbance measurements. With added injectors, the Victor 2 performs fill, empty or flush operations to designated plate.

A bench top unit, VICTOR2 can operate as a stand-alone instrument; some models can be prepared with input and output stackers and bottom reading optics. Some models include injectors for fluorescence and flash luminescence.

 

Fluorescence Range (340 – 850 nm)
• Dual ratio measurements
• Top down readings fluorescence
Luminescence features
• Glow luminescence
• Flash luminescence use with injector only
• Dual luminescence

Photometry Range (340 – 800 nm)

 

Microwell Plates 1-1536 wells (luminescence up to 384 wells)
Dispense injector (specified models) 5-350µl in 5µl increments
Scanning over wells and kinetics Yes, included with software
Shaking features Linear, orbital or double orbital
Absorbance Filters in an 8 place filter wheel Installed 405, 450nm filters or user choice
Fluorescence Filters in an 8 place filter wheel and interchangeable slide  
Excitation Filters installed in an 8 place filter wheel or choice 2 pairs

355, 485nm or user choice

Emission filters installed in 8 position filter slide or choice 2 pairs

460, 535nm or user choice

Fluorometry standards specifications Filters are high quality with 15nm bandwidth
Fluorescein < 2 fmol/well
Linearity > 5 decades
Crosstalk < 0.0001
Umbelliferone < 200 fmol/well
Rhodamine < 100 fmol/well
Photometry standards specifications  
Measuring range @ 405 nm 0 – 2 A
Accuracy @ 405 nm < 2% or (0.01A)
Precision @ 405 nm < 0.5% or (0.01A)
Luminometry standards specifications  
Lower limit of detection < total flux of 105 photons/sec
ALP typically 1 amol/well with AMPPD substrate
ATP typically 80 amol/well in flash assays
Luciferase 0.9 pg/well using Genglow-100 or GeneLux
Ordering Information:  
Victor 2 stand alone Catalog #P97095 (1 in stock)
Victor 2 with 1 injector module Catalog #P97092 (1 in stock)
Victor 2 with 2 injector modules Catalog #P97093 (1 in stock)

National Institutes Standards
& Technology Plate

Catalog #9850015

Warranty information:

Comes complete with line cord, PC software, NIST documentation and certification, one year warranty and six days a week technical support.