Microplate Instrumentation Specialists

          Ordering Information
          Phone   1.800.848.6474   
          Phone 01.703.821.1045
          Fax      01.703.821.1046
          Email  
info@mtxlsi.com

Additional  in-stock Fluorescence Plate Readers:
|Cytofluor 4000|BMG Fluostar |BMG Polarstar|Bio-tek FL 600|

|Tecan Spectrafluor |Tecan Spectrafluor Plus |

|Molecular Devices Gemini EM|

|Perkin Elmer Victor 2 |Perkin Elmer Victor|Perkin Elmer HTS 7000|

|Thermo Electron/Labsystems Finstruments Fluorescence|

|Thermo Electron Ascent CF|Fluoroskan II|
Fluoroskan I|Accessories|


Perkin Elmer Victor3v model
Fluorescence, Absorbance and Luminescence Reader  SOLD
 

Description:

VICTOR3V is a top of the line plate reader for all detection technologies including: fluorescence (top and bottom), luminescence, absorbance, UV absorbance, time-resolved fluorometry, and fluorescence polarization technologies.

A small bench top unit, VICTOR3V can operate as a stand-alone instrument or can integrate into a robotic system. Outboard injectors module for solid flash luminescence performance can be added.


Fluorescence (340 – 850 nm)
• Dual ratio measurements
• Top and bottom fluorescence
Luminescence
• Glow luminescence
• Flash luminescence
• Dual luminescence

Photometry (340 – 800 nm)
UV Photometry (260 nm/280 nm)
Time-resolved Fluorometry (TRF)
• Dual window TRF
• Dual emission measurements
Fluorescence Polarization:
(400 – 850 nm)

A wide range of high quality filters is factory installed. These filters cover the most common wavelengths used in fluorometric and photometric applications. Additional filters are available upon request.

Absorbance 405 nm, 450 nm and 492 nm
UV absorbance
260 nm and 280 nm
 
Fluorescence  
Excitation

355 nm and 485 nm

Emission

460 nm and 535 nm

TR Fluorescence

 

Excitation

340 nm

Emission

545 nm, 572 nm, 616 nm and 642 nm

Additional 665 nm emission filter for LANCE™ comes with red sensitive PMT

 
Fluorescence polarization  
Excitation 480 nm and 531 nm
Emission 535 nm and 595 nm
Fluorometry  
Fluorescein < 2 fmol/well
Linearity > 5 decades
Crosstalk < 0.0001
Umbelliferone < 200 fmol/well
Rhodamine < 100 fmol/well
Time Resolved-Fluorometry  
Eu typically < 10 amol/well
Eu linearity > 5 decades
Crosstalk < 0.0001
Tb typically < 5 amol/well
Sm typically < 50 amol/well
Dy typically < 150 amol/well
Photometry 340-800nm
Measuring range @ 405 nm 0 – 2 A
Accuracy @ 405 nm < 2% or (0.01A)
Precision @ 405 nm < 0.5% or (0.01A)
Luminometry  
Lower limit of detection < total flux of 105 photons/sec
ALP typically 1 amol/well with AMPPD substrate
ATP typically 80 amol/well in flash assays
Luciferase 0.9 pg/well using Genglow-100 or GeneLux
Fluorescence Polarization 400-850nm
Fluorescein 1 nM (384-well plate)
Standard deviation < 5 mP
   
Ordering Information  

Victor 2

Catalog #P97090
Victor 3v NEW Catalog #P97091

Warranty information:

Comes complete with line cord, PC software and six days a week technical support. New model with one year warranty.

Microplate Instrumentation Sales, Service and Support since 1982.