Description:
VICTOR3V is a top of the line plate reader for all detection
technologies including: fluorescence (top and bottom), luminescence,
absorbance, UV absorbance, time-resolved fluorometry, and
fluorescence polarization technologies.
A small bench top unit, VICTOR3V
can operate as a stand-alone instrument or can integrate into a
robotic system. Outboard injectors module for solid flash
luminescence performance can be added.
|
Fluorescence
(340 850 nm)
Dual ratio measurements
Top and bottom fluorescence
Luminescence
Glow luminescence
Flash luminescence
Dual luminescence |
Photometry (340
800 nm)
UV Photometry (260 nm/280 nm)
Time-resolved Fluorometry (TRF)
Dual window TRF
Dual emission measurements
Fluorescence Polarization:
(400 850 nm) |
A wide range of high quality filters is factory
installed.
These filters cover the most common wavelengths used
in fluorometric and photometric applications. Additional
filters are available upon request.
|
Absorbance |
405 nm, 450 nm and 492 nm |
|
UV absorbance |
260 nm and 280 nm |
|
Fluorescence |
|
|
Excitation |
355 nm and 485 nm |
|
Emission |
460 nm and 535 nm |
|
TR Fluorescence |
|
|
Excitation |
340 nm |
|
Emission |
545 nm, 572 nm, 616 nm and 642 nm |
|
Additional 665 nm emission filter for
LANCE
comes with red sensitive PMT |
|
|
Fluorescence polarization |
|
|
Excitation |
480 nm and 531 nm |
|
Emission |
535 nm and 595 nm |
|
Fluorometry |
|
|
Fluorescein |
< 2 fmol/well |
|
Linearity |
> 5 decades |
|
Crosstalk |
< 0.0001 |
|
Umbelliferone |
< 200 fmol/well |
|
Rhodamine |
< 100 fmol/well |
|
Time Resolved-Fluorometry |
|
|
Eu typically |
< 10 amol/well |
|
Eu linearity |
> 5 decades |
|
Crosstalk |
< 0.0001 |
|
Tb typically |
< 5 amol/well |
|
Sm typically |
< 50 amol/well |
|
Dy typically |
< 150 amol/well |
|
Photometry |
340-800nm |
|
Measuring range |
@ 405 nm 0 2 A |
|
Accuracy |
@ 405 nm < 2% or (0.01A) |
|
Precision |
@ 405 nm < 0.5% or (0.01A) |
|
Luminometry |
|
|
Lower limit of detection |
< total flux of 105 photons/sec |
|
ALP typically |
1 amol/well with AMPPD substrate |
|
ATP typically |
80 amol/well in flash assays |
|
Luciferase |
0.9 pg/well using Genglow-100 or GeneLux |
|
Fluorescence Polarization |
400-850nm |
|
Fluorescein |
1 nM (384-well plate) |
|
Standard deviation |
< 5 mP |
| |
|
| Ordering
Information |
|
|
Victor 2 |
Catalog #P97090 |
| Victor
3v NEW |
Catalog #P97091 |
|
Warranty information:
Comes complete with line cord,
PC software and six days a week technical support. New model
with one year warranty.
Microplate Instrumentation Sales, Service and Support since 1982.
|