TECAN Genios Pro Multifunction Microplate Reader Fluorescence TRF, FI, FRET
Absorbance
and Luminescence Microplate Reader
 for top- and bottom- read assays

All new filters, parts, lamps and 2 year warranty!



 

Instrument Description and Capabilities:  

The Tecan Genios Pro is a fully automatic, computer controlled Fluorescence. Luminescence and Absorbance instrument for measuring samples in a microplate. The instrument fulfills the requirements for numerous fluorescence, absorbance and luminescence applications in research and routine investigations.

The filters are arranged in filter slides; four excitation filters and four emission filters respectively can be installed in the slides. Handles 6, 24, 96 and 384 well plates. Spectral range 360-700nm for fluorescence. 230-900nm absorbance spectral range. Sensitivity 1.5 picogram fluorescein on each well or ≤4 fmol well. The emission signal is measured by a highly sensitive PMT. Different detection techniques with one instrument change of measurement directions (top / bottom) by PC software.  Thermal incubation at  +5º C to + 40º C.

Dual optical channels allow measurements from above or below the plate by mouse-click. Light to and from the samples is focused by a lens and high quality filters are easily accessible from the front of the instrument. Each filter slide can hold up to four filters.
Orbital shaking and temperature control are standard.

Product Specifications include:

  • Fluorescence (VIS range)
  • Time Resolved Fluorescence TRF
  • Fluorescence Intensity top and bottom capabilities (UV to NIR) FI
  • Fluorescence Resonanace Energy Transfer FRET
  • Homogeneous Time Resolved Fluorescence (HTRF®)
  • Absorbance reading mode (UV to NIR range)
  • Temperature control and microplate-shaking included
  • Adjustable bandwidth range
  • Luminescence Flash reading mode
  • Dual Color Luminescence
  • 6-384 well plate capability
  • PC Computer with Magellan Software
  • Advanced injector design allows simultaneous dispensing & reading
  • Additional injectors can be used in 6-384 well plates & all reading modes
  • High-quality syringe pumps guarantee high precision & optimum mixing
  • Bump-back function minmizes dead-volume and conserves valuable reagents
  • High-end mirror optics and three detectors guarantee maximum sensitivity
  • Well-scanning function available for compensation of uneven cell distribution
  • Multi-label kinetic reading provides higher information content per application
  • Modular approach allows easy adaptation to changing application needs
  • XFluor™ software for direct control from Excel included
  • Magellan™ software additionally offers full data reduction
  • Advanced QC-tools and IQ OQ documents available

 

Plate formats 6 to 384 well plates
Absorbance Range 230-999nm
Temperature control
Ambient + 5 °C to 42 °C
 
Shaking linear or orbital, 3 velocities
Fluorescence Intensity

Wavelength range

Excitation: 230 – 850 nm, emission:
280 – 850 nm

Bandwidth

Adjustable from 2.5 nm – 10 nm below 280 nm, 5 nm – 20 nm else

Wavelength accuracy

± 1 nm

Wavelength precision

± 1 nm
Sensitivity* – top reading  
– 384 well plate 0.2 fmol / well Fluorescein
Sensitivity* – bottom reading  
– 384 well plate 0.6 fmol / well Fluorescein
Time Resolved Fluorescence  
Sensitivity* – 384 well plate 60 amol / well Europium
Precision – 384 well plate 3 mP standard deviation @ 1nM Fluorescein
Fluorescence Polarization Wavelength Range 275-750nm
Absorbance 96 well  
Wavelength range 230 – 1000 nm
Wavelength accuracy 492NM ±/- 1% and ±/- 0.010 OD OD
Wavelength precision 492NM ±/- 10,5% and ±/- 0.005 OD OD
Measurement range 0 – 4 OD
Glow Type Luminescence Chemi or Bioluminescence
Wavelength readings 400-700NM
Read times – fast scanning
Scanning of 96 well plate, 35 sec for FI, TRF and Absorbance
Scanning of 96 well plate, 70 sec for FP
Scanning of 384 well plate, 65 sec for FI, TRF
Scanning of 384 well plate, 85 sec for FP
Interface Connections RS 232 serial & parallel port
Tecan Genios Pro Ordering Information Catalog #P97084 Remanufactured  (3)

Dimensions: LxWxH: 48 x 44 x 27 cm; weight 28 kg
Warranty information:

Instruments come complete with NIST certification and documentation on filters, line cord, PC software and six days a week technical support. Remanufactured model with  warranty.  High quality instrument made in Austria and Germany. Easily reset to 220 Volt.

Microplate Instrumentation Sales, Service and Support since 1982.