| Instrument Description
and Capabilities:
TECAN ULTRA Microplate
Reader, a
multi-functional microplate reader developed especially for life
scientists working with very high throughputs. The ULTRA lets you
choose between several fluorescence techniques as well as absorbance
and luminescence modes, without having to change the optical system.
Specifications include:
- Fully modular system
- Top and bottom Fluorescence
Intensity capabilities FI
- Excitation-, Emission-,
Fluorescence
- Fluorescence Resonanace Energy
Transfer FRET
- Adjustable bandwidth
- Fluorescence Time Resolved
TRF
- Fluorescence Polarization FP
- Multichannel Absorbance
- Multichannel Luminescence Glow
Type
- 6-1536-well plate capability
- Fast wavelength scanning
capabilities
- High sensitivity for short
measurement times
With years of TECAN expertise behind it, the ULTRA gives you precise
readings every time, even at very low detection limits. This flexible
reader can handle any plate format up to 1536 wells, for all
techniques. It can read a 1536 well plate in less than a minute and so
is ideal for very high throughput applications when you need results
immediately.
Three axis positioning (x, y, z) allows you to select the best
measuring conditions to suit different well shapes, sample volumes and
assay types, giving outstanding sensitivity. Various filters are
available for different applications, including special dichroic
mirrors for Fluorescein, Europium and Cy5 dyes.
The ULTRA is a fully automatic, computer controlled
Fluorescence, Fluorescence Intensity,
FRET, TRF, FP, Luminescence and Absorbance instrument for measuring samples in a microplate. The instrument
fulfills the requirements for numerous fluorescence, luminescence and absorbance
applications in research and routine investigations.
The filters are arranged in filter slides; four
excitation filters and four emission filters respectively can be
installed in the slides. Handles 6, 24, 96, 384
and 1536
well plates. Spectral range 230-999nm for fluorescence. 230-1000nm
absorbance spectral range. Sensitivity ≤1.5 picogram fluorescein on each well or ≤4 fmol well. The
emission signal is measured by a highly sensitive PMT. Different
detection techniques with one instrument change of measurement
directions (top / bottom) by PC software. Thermal incubation at +5º C to +
42º C.
Dual optical channels allow measurements from above or below the plate
by mouse-click. Light to and from the samples is focused by a lens and
high quality filters are easily accessible
from the front of the instrument. Each filter slide can hold up to four
filters.
Orbital shaking and temperature control
are standard.
Suggested
use with black plates for bottom-to-top readings.
Optimal use with black plates prevent cross-contamination.
| Plate formats |
6 to 1536 well plates |
| Absorbance Range |
230-999nm |
| Temperature
control |
Ambient + 5 °C to 42 °C
|
| Shaking |
linear or orbital, 3 velocities |
| Fluorescence Intensity |
|
| Wavelength range |
Excitation: 230 – 850 nm,
emission: 280 – 850 nm |
| Bandwidth |
Adjustable from 2.5 nm – 10 nm
below 280 nm,
5 nm – 20 nm else |
| Wavelength
accuracy |
± 1 nm |
Wavelength precision |
± 1 nm |
| Sensitivity* –
top reading |
|
| – 384 well plate |
0.2 fmol / well Fluorescein |
| Sensitivity* – bottom reading |
|
| – 384 well plate |
0.6 fmol / well Fluorescein |
| Time Resolved Fluorescence |
|
| Sensitivity* – 384
well plate |
60 amol / well Europium |
| Fluorescence Polarization |
|
| Precision – 384 well
plate |
3 mP standard deviation @ 1nM
Fluorescein |
| Fluorescence Polarization |
Wavelength Range 275-750nm |
| Absorbance |
|
| Wavelength range |
230 – 1000 nm |
| Bandwidth |
2.5 nm below 280nm, 5 nm else |
| Wavelength
accuracy |
± 1 nm |
| Wavelength precision |
± 1 nm |
| Measurement range |
0 – 4OD |
| Accuracy @ 492 nm |
0 – 2OD: ± 1% and 0.006OD |
| Precision @
492 nm |
0 – 2OD: ±
0.5% and 0.005OD |
| Glow Type Luminescence |
Chemi or Bioluminescence |
| Wavelength
readings |
400-700NM |
| Fluorescence Absorbance FP |
|
| 384 well plate |
25 sec 25 sec 45
sec |
| 1536 well
plate |
37 sec 27 sec 108
sec |
| Read times – fast scanning |
Fluorescence
scanning of 384 well plate, 450 nm – 550 nm,
5nm step: 321 sec for 384 spectra |
Absorbance scanning of 384 well plate, 450 nm – 550 nm,
5nm step: 103 sec for 384 spectra |
| Interface Connections |
RS 232 serial
& parallel port |
| Tecan
Ultra Ordering Information |
Catalog #P97087 Remanufactured (1) |
|