POLARstar - Technical Specifications |
Measurement Principles |
Fluorescence
Fluorescence Intensity
Time-Resolved Fluorescence
Polarization (some models)
Luminescence (some models)
Absorbance |
Light Source |
High-energy xenon flash lamp |
Detector |
Side window, current type photomultiplier tube |
| Fluorescence Wavelength Range |
240-740nm spectral range |
Fluorescence Filters |
Excitation and Emission filter wheels
accommodate 8 and
4 filters respectively
with your choice of 2 filter pairs installed in instrument |
| Absorbance Filters |
One
filter wheel accommodate 4 filters each
with your choice of 1 filter installed in instrument |
Fluorescence Intensity |
Limit of detection: 0.9 fmol / well fluorescein
calculated according to the IUPAC standard: 3(SDblank)/slope |
Time-Resolved Fluorescence |
Limit of detection: 50 amol/well europium
calculated according to the IUPAC standard: 3(SDblank)/slope |
Absorbance
Wavelength Range |
240-740nm
spectral range
Dynamic range: 0.000-3.000 OD
Reproducibility: 0.010 OD for 0-2 OD range |
Microplate Formats |
up to 384 well and
Fluorescence intensity and time-resolved fluorescence up to 1536 well
plate.
Fluorescence black plate and Absorbance clear plate |
Reading Time |
Depends on test definition
shortest reading time: 15 s for 96-well plate
30s for 384 well plate 60s for 1536 well plate |
Incubation |
Incubation range from ambient +8 °C to 45 °C
in 0.1°C steps
Temperature stability 0.2 °C
Extended incubation up to 60°C |
Shaking |
Orbital or linear shaking, user definable time and
shake width |
|
Reagent injectors |
Speed 100µl/s to 420µl/s
Volumes down to 3µl and for each well
Up to 4 independent injection actions per well |
Dimensions/weight |
LxWxH: 48 x 44 x 27 cm; weight 28 kg |
|
Interface
connections |
RS 232 serial & parallel
port |
Catalog #P21007 POLARstar Galaxy with 2 injectors
Catalog #P21008 POLARstar
Optima
Remanufactured with one year warranty. |