Additional in-stock Multi-mode Detection
and Fluorescence Plate Readers:
Click on link to go to page
|Fluoroskan I|Perseptive Biosystems Cytofluor 4000|BMG
FLUOstar|
|Biotek
FL 600|Dynex
Technologies MLX|
|ThermoFisher Finstruments Fluorescence|ThermoFisher
Ascent|
|Tecan
Spectrafluor|Tecan
Spectrafluor Plus|Tecan
Genios|
|Molecular
Devices Gemini XS|ThermoFisher
Luminoskan|
|ThermoFisher
Luminoskan Ascent|
|BMG
FLUOstar Galaxy|Perkin
Elmer Victor 2 with or w/o injectors|
|BMG POLARstar|
|BMG POLARstar
Galaxy|Biotek
FLX 800|BMG
NOVOstar|Tecan
Ultra|
|Tecan Genios Pro|Thermo Fisher
Ascent FL with w/o injectors|
|Molecular Devices
M5|Molecular
Devices Gemini EM|Tecan
Safire|
Dynex Technologies MLX 96 Well Plate
Luminometer Equipped Luminescence Reader
for top- and bottom- read assays from
$5,995.
The
MLX, a highly sensitive microplate luminometer with on-board computer, uses
the familiar Revelation software platform for data analysis and reporting.
With excellent sensitivity, eight decades of dynamic range and a specially
designed plate mask to minimize cross talk, the MLX Luminometer helps you
produce consistent results plate-to-plate, day-to-day and
instrument-to-instrument. This model comes with 0-2 positive displacement
dispensers and temperature control to 45 *C. (as noted)~ Sensitivity: Alkaline phosphatase (10-21 moles) ~ Dynamic range: 0.0008 - 10,000 RLU ~ Linearity:
R2 > 0.999 from 0.001 - 9900 RLU ~ Precision: Better than 1% from 0.3 - 9900
RLU ~ Cross talk: <2.5 x 10-5 from 0.0001 - 999.99 RLU.
The MLX Microtiter
Luminescence Detection System is highly sensitive and geared toward
measuring luminescence in 96-well plates.
This instrument contains Revelation,
Dynex's operating and data analysis software in Windows format. This
luminometer features a broad temperature control range, from 5°-45°C. Long
Dynamic range at 8 decades 0.0008-10,000. Monitor, Keyboard, Mouse not included.
PC Software is contained on
the hard drive.
Instrument
Description and Capabilities:
Long Dynamic Range
- 8 Decade
- High, Medium, Low gain
settings
- Auto gain for Flash and
Glow reactions
Light Sampling Rate -
100 Readings per Second
- Highest resolution for
flash reactions
- Better reproducibility
with more readings per user set time
- Adjustable by User
through software
Unique and Patented
Mask Design
- Part of plate carrier
- Ensures minimal
cross-talk
- Self-adjustment of plate
against mask using locking elevator
Photomultiplier Tube
- Self-protected with
shutter
- Shutter creates a light
tight environment enclosing the PMT and LED
- Shutter allows
calibration for dark current or electronic noise of the PMT
Internal LED
- Allows calibration for
drift, PMT ageing
- Has photodiode to check
light output
Dispensing
- 0-3 Dispensers on Board
(optional)
- 500 µl total volume
syringes
- Simultaneous or
sequential dispensing programmable
- Inert, PEEK or PTFE
materials only which reagents touch
- Dark cover protection of
fluids from light degradation
- Expensive reagent
recovery by programmable "back pedaling"
Information Exchange
- Files can be
generated to send to Excel
Temperature
Control
- Base plate is temperature
controlled
- Ambient plus 5°C - 45°C
Analytical Performance
- Sensitivity
> 10 fentomoles of alkaline phosphatese
1 x 10-18 moles of aequorin
- Precision
>1% from 0.1 - 9900 RLU
- Linearity
r2>0.999 from 0.001 to 9900
- Read Time
30 seconds chemistry dependent
- Cross Talk
< 2.5 parts per 100,000
Flexible Operating
Programs (Revelation
PC Software)
- Glow end-point
Alkaline phosphatase dioxetane substrates
B-galactosidase
Flow Kinetic Multi-Cycle
Flash
Luciferase
Aequorins
Acridinium esters
Linked Assays
User designed processing routines
Data Reduction
- Quantitative Analysis
Flexible Templates, Multiple curve fits
Multiple assays per plate, Any blanking
- Qualitative Analysis
Multiple ratios, differencing, spreadsheets, threshold screening
Assay Password Protection Assay Wizards
- Intuitive guide to
programming assay setups
- Saves programming time
- Ensures no important
steps are forgotten
Quick Start Buttons
- Press on button to start
reading a plate
- Efficient day-to-day use
- Operator-to-operator same
assay
Auto Save Data
- Data saved automatically
following plate read
- Allocated sequential
numbers for multiple plate reads
- User recallable for
calculation manipulations
- Graphs also saved for
recalculation later
User needs monitor, keyboard, mouse and printer for operation.
Catalog # P300011 |
Dynex MLX Luminometer
no dispensers
Remanufactured (2) |
|
Catalog # P300012 |
Dynex MLX Luminometer
1 dispenser
Remanufactured (2) |
Catalog #
P300013 |
Dynex MLX Luminometer
2 dispensers
Remanufactured (2) |
Warranty information:
Comes complete with 110 Voltage line cord
(easily reset to 220 voltage),
manual, technical support and warranty! Highest quality instrument.
Easy operation with PC software, accurate and precision
oriented reader.
Microplate Instrumentation
Sales, service and support since 1982.
Several always in stock,
recently serviced and documented in perfect condition |