Technical Specifications
|
Measurement Principles |
Fluorescence
Fluorescence Intensity
Time-Resolved Fluorescence
Absorbance
Luminescence |
Light Source |
High-energy xenon flash lamp |
Detector |
Side window, current type photomultiplier tube |
Filters |
Two filter wheels for 8 filters each
with your choice of 2 filter pairs of choice |
Gain Control |
Automatic gain adjustment |
Fluorescence Intensity |
Limit of detection: 0.9 fmol / well fluorescein
calculated according to the IUPAC standard: 3(SDblank)/slope |
Time Resolved Fluorescence |
Limit of detection: 50 amol/well europium
calculated according to the IUPAC standard: 3(SDblank)/slope |
Absorbance |
Dynamic range: 0.000-3.000 OD
Reproducibility: 0.010 OD for 0-2 OD range |
Microplate Formats |
up to 96 well
Fluorescence intensity and time-resolved fluorescence up to 1536 well |
Reading Time |
Depends on test definition
shortest reading time: 15 s for 96-well plate |
Incubation |
Incubation range from ambient +5 °C to 45 °C (option: 60 °C)
Temperature stability 0.2 °C
Temperature gradient < 0.5 °C |
Shaking |
Orbital or linear shaking, user definable time and amplitude |
| Interface Connections |
RS 232 serial
& parallel port |
|
Dimensions/Weight |
LxWxH: 48 x 44 x 27 cm; weight 28 kg |
|
Catalog # P21010 |
FLUORstar Model
Remanufactured |
|
Catalog # P21012 |
FLUORstar
Model upgraded Remanufactured |
|
Catalog # P21004 |
FLUORstar Galaxy with 2
injectors Remanufactured |
|
Catalog # P21006 |
POLARstar Remanufactured |
|
Catalog # P21009 |
POLARstar Galaxy
Remanufactured
|
|
Catalog # P21007 |
POLARstar Galaxy
with 2 injectors Remanufactured
|